ORIGINAL ARTICLE
Year : 2018  |  Volume : 30  |  Issue : 1  |  Page : 62-65

Evaluation of bacterial contamination of dental unit water lines before and after the use of 1% (Lysoformin 3000) disinfectant and awareness of the dental unit waterline disinfection among the practicing dental surgeons of Vadodara city


Department of Conservative Dentistry and Endodontics, K.M. Shah Dental College and Hospital, Sumandeep Vidyapeeth, Vadodara, Gujarat, India

Date of Web Publication4-Jun-2018

Correspondence Address:
Dr. Nimisha Shah
Department of Conservative Dentistry and Endodontics, K.M. Shah Dental College and Hospital, Sumandeep Vidyapeeth, At & Po Pipariya, Waghodia, Vadodara, Gujarat
India
Login to access the Email id

Source of Support: None, Conflict of Interest: None


DOI: 10.4103/endo.endo_47_17

Rights and Permissions
  Abstract 

Purpose: The aim is to evaluate the bacterial contamination in dental unit waterlines (DUWLs) before and after the use of the disinfectant 1% (Lysoformin 3000) for its efficacy also assess the awareness about the same among the practicing dental surgeons of Vadodara city.
Methodology: A volume of 20 ml water samples from the DUWLs from the clinics of the practicing dental surgeons of Vadodara city was collected. And were asked to answer a validated questionnaire in private. Disinfection of the waterlines was performed twice a week for 2 weeks with 1% disinfectant (Lysoformin 3000) and the samples were again collected. Qualitative and quantitative microbiological analysis was performed for both the predisinfected and postdisinfected samples.
Results: Predisinfection samples showed a mean colony count of 670.35 colony forming unit (CFU)/ml which was reduced to 63.90 CFU/ml (P < 0.001) postdisinfection. Questionnaire assessment showed that 70% awareness of the participants about DUWL infections and the microorganisms associated. However, only 20% practiced any type of DUWL disinfection.
Conclusion: Lysoformin 3000 can be promising to be used a DUWL disinfection solution as an easy to use and readily available solution with great antibacterial properties. However, still more long-term studies are required to check for its efficacy in anaerobic bacteria as well as resistant species of microorganisms.

Keywords: 1% Lysoformin 3000, biofilm, dental unit waterline


How to cite this article:
Bhadra D, Shah N, Patel P, Dedania M. Evaluation of bacterial contamination of dental unit water lines before and after the use of 1% (Lysoformin 3000) disinfectant and awareness of the dental unit waterline disinfection among the practicing dental surgeons of Vadodara city. Endodontology 2018;30:62-5

How to cite this URL:
Bhadra D, Shah N, Patel P, Dedania M. Evaluation of bacterial contamination of dental unit water lines before and after the use of 1% (Lysoformin 3000) disinfectant and awareness of the dental unit waterline disinfection among the practicing dental surgeons of Vadodara city. Endodontology [serial online] 2018 [cited 2018 Jun 22];30:62-5. Available from: http://www.endodontologyonweb.org/text.asp?2018/30/1/62/233738


  Introduction Top


Biofilm contamination in dental unit waterlines (DUWLs) remains a considerable problem in dentistry's as the reported bacteria levels in water from DUWLs can often go over the limit exceeding 100,000 colony forming units (CFUs)/ml of water, over 200 times the amount of allowable bacteria in potable water.[1] DUWLs in dentistry are used to cool and irrigate tooth surfaces during treatment as well as dental handpieces.[2] They includes narrow-bore, plastic tubing, which are highly susceptible to biofilm contamination due to reduced velocity of water at the periphery of the tube.[2] The Environmental Protection Agency, American Public Health Association, and American Water Works Association have all set limits for heterotrophic bacteria of <500 CFU/ml.[3] The Centers for Disease Control and Prevention published “Guidelines for infection control in dental health care settings” in 2003, and highlighted that oral Streptococcus species, human pathogens such as Pseudomonas aeruginosa and Legionella species, and nontuberculosis Mycobacterium species had all been isolated from dental water systems.[3],[4],[5] In numerous studies, cross contamination between patients has been reported.[6],[7],[8] Suck-back of oral fluids from a patient, either from protective anti-retraction valve failure, or when handpieces are removed and replaced incorrectly have been the reason for this contamination.[7],[8] The most efficient means of maintaining good quality dental chair unit (DCU) output water is regular disinfection of DUWLs with a disinfectant or biocide that removes biofilm from the water lines resulting in output water of potable quality.[9] A wide variety of commercial waterline cleaning products and systems are available, some of which can be retrofitted to existing DCUs. One such disinfecting agent is Lysoformin 3000, which has an excellent bactericidal and bacteriostatic activity. It is used in various concentrations ranging from 1% to 1.5%. Among which 1% disinfectant (Lysoformin 3000) is a good antiseptic that has been used for disinfection of medical apparatus since a long time and can be used as a DUWL disinfectant. However, very few studies published in the peer-reviewed literature have actually investigated the efficacy of this disinfectant to achieve these desired effects in DCUs and detailed comparative studies have yet to be published.[9] Hence, there is a need to study the efficacy of this disinfectant in reducing the microbial load in DUWL. Furthermore, there is an onus on DCU manufacturers to consider the problem of DUWL biofilm contamination when designing DCUs. Hence the aim the study is to evaluate the awareness of the practicing regarding disinfection of DUWLs and to check the efficacy of 1% disinfectant (Lysoformin 3000) in disinfection of DUWLs. The null hypothesis of the study was that there will be no reduction in the load of microorganisms in the DUWL after the use of 1% (Lysoformin 3000) disinfectant.


  Methodology Top


Twenty practicing dental surgeons practicing in Vadodara city were selected for the study by lottery method from the Indian Dental Association directory. Dental surgeons having their private practice in the city and consenting for the study were included in the study. All the participants were asked to fill a questionnaire that was content validated and subject validated at 86% (Chi-square test) in private revealing their awareness regarding DUWL disinfection. Twenty milliliters predisinfection water samples were then collected in sterile plastic containers (ABDOS P40102) from the three-way syringe of the dental chair. Moreover, the subjected to microbiological evaluation. One percent disinfectant (Lysoformin 3000) was then added to bottle booster system in the dental chair and passed through the waterline into the three-way syringe till a blue colored liquid was seen. It was then allowed to stay in the waterline overnight and then washed away from the waterline and flushed by distilled water. This procedure was repeated twice a week for 15 days. Then, postdisinfection water samples were collected and also subjected to microbiological analysis.

Microbiological evaluation

Water samples were then centrifuged at 3000 rpm and the supernatant was discarded, and the centrifuged sample was cultured in a MacConkeys agar without crystal violet (high media) at 33°C for 48 h followed by colony counting. Bacterial identification kit was used for the identification of the bacteria present in the culture.

This data were then subjected to statistical analysis.


  Results Top


Paired sample t-test showed a mean colony count of 670.35 CFU/ml before disinfection which was reduced to the mean of 63.90 CFU/ml (P< 0.001) [Table 1] and [Graph 1] which was highly statistically significant. Bacterial identification confirmed the presence of  Escherichia More Details coli, Staphylococcus aureus, Klebsiella pneumoniae and P. aeruginosa and postdisinfection left only the traces of E. coli and S. aureus.
Table 1: Paired t-test showing mean colony forming unit count pre- and post-disinfection of dental unit waterlines

Click here to view



Questionnaire assessment [Table 2] of the validated questionnaire 86% (Chi-square test) showed that 70% of the participants were aware about DUWL infections and the microorganisms associated. However, only 40% knew the evidence-based methods and products regarding DUWL disinfection, whereas only 20% practiced any type of DUWL disinfection protocol regularly and sodium hypochlorite was the material of choice for all.
Table 2: Validated questionnaire showing responses of the participants regarding dental unit waterlines disinfection

Click here to view



  Discussion Top


The result of this study showed a statistically significant reduction in the number of CFUs postdisinfection with the help of 1% disinfectant (Lysoformin 3000). Such major differences in the results obtained are undoubtedly related to the composition of Lysoformin 3000, where, although the main active substance is glutaraldehyde, there are also other compounds considered disinfectants, such as didecyldimethyl ammonium chloride, isotridecanolethoxylated, and glyoxal.[10] Due to the drawbacks and limitations in disinfection of general disinfectants, it is believed that multi-component disinfectants are the optimal solution, as they combine the action of all the ingredients and enhance the properties.[10] The combination of gluteraldehyde activity due to the alkaline reactions and the quaternary ammonium compounds in Lysoformin 3000 mark for its antibacterial as well as antifungal effectivity.[11] Likewise, studies have reported other disinfectants such as Tegodor and Gigasept Rapid (aldehyde based) achieved a 100% reduction in the biofilm viability but did not remove the biofilm adhered to the tubing surface. This phenomenon is mostly associated with alkaline reactions of glutaraldehyde that are responsible for cell wall lysis of the bacteria rendering the compound to be more effective.[11]

The presence of quaternary ammonium compounds summated the efficacy of the disinfectant (Lysoformin 3000). The high efficacy of disinfectant substances tested containing quarternary ammonium salts (QAT) corresponded with results of antibacterial activity of homologous series of alkyl trimethyl ammonium bromides and dialkyl dimethyl ammonium bromides against the clinical  Salmonella More Details typhimurium and P. aeruginosa strains,[12] and one important factor of QAT efficacy was found the length of the alkyl chain.[12],[13] Similar results were obtained with E. coli strains.[14] QAT belong to the membrane active compounds binding to the cytoplasmic membrane and influencing cell metabolism.[15] Quaternary ammonium compounds gained popularity due to its good antibacterial and antifungal at relatively low concentration within shorter contact time.[16] Moreover, the blue color of the solution and a typical odor makes it easy, comfortable and safe for usage as well as cleaning.[17]

However, there has been a varied range of disinfection modalities right from the periodic replacement of the waterlines or autoclaving of the waterline units, anti-retraction valves, etc. However, these seem to bit more tedious and non-user-friendly in developing country like ours. Hence, the disinfecting solutions and tablets can make a very comfortable and effective choice for DUWL disinfection for the dental offices in India. Lysoformin 3000, is highly effective against both Gram-positive and Gram-negative bacteria at all concentrations and contact time tested the aldehyde-based disinfectant (Lysoformin 3000). These findings are similar to another study, which showed that 2% Lysoformin 3000 at 30 min was enough for bactericidal and fungicidal activity.[18] Lysoformin 3000 was found to be effective against bacterial spores as reported.[18],[19]

Staniszewska et al. stated that glutaralaldehyde in the concentration of 1%–2% reduces Bacillus cereus spores within 15–30 min by 4 orders of magnitude. The results of the study carried out were different, and demonstrated that Lysoformin 3000 with glutaraldehyde as its main active substance reduces totally the spores of all three strains, namely B. subtle combine the active component with components that have washing properties or are emulgators.[20]Bacillus mycoides, and B. cereus already after 5-min exposure. Hence, 1% disinfectant (Lysoformin 3000) was used for the disinfection.[21]

This was followed twice weekly for 15 days. The disinfection protocol was designed accordingly because, In a controlled study, twice a week overnight (15 h) disinfection using either agent reduced the bacterial density to below the American Dental Association recommended the level of 200 CFU/ml.[9]

Questionnaire assessment showed that 70% to the participants were aware about DUWL infections and the microorganisms associated. However, only 40% knew the evidence-based methods and products regarding DUWL disinfection, while only 20% practiced any DUWL disinfection protocol regularly and sodium hypochlorite was the material of choice for all.

Hence, the need of the hour is a readily available, easy to use, noncaustic disinfectant that will be effective against a wide range of microorganisms associated with the hospital-acquired infection in a dental office. Lysoformin 3000 can very well be the hero to cometh the moment. However, still more long-term studies are required to check for its efficacy in anaerobic bacteria as well as resistant species of microorganisms.

However, a greater sample size would have given us a better judgment regarding the contamination of DUWL and the awareness about the same among the fellow practicing dental surgeons. A longer and continuous use of disinfectant 1% (Lysoformin 3000) would have enlightened us regarding the long-term efficacy of the disinfectant.

As medical professionals 0% infection is our goal. Hence, fire must be ignited always. Quest for the best must not end.


  Conclusion Top


Within the limitations of this study, 1% disinfectant (Lysoformin 3000) can be promising to be used as a DUWL disinfection solution as an easy to use and readily available solution with great antibacterial properties. However, more long-term studies are required to check for its efficacy in anaerobic bacteria as well as resistant species of microorganisms. Furthermore, there seems to be a need to educate and guide regarding the DUWL disinfection in practicing dentists as well as at the undergraduate and postgraduate levels.

Financial support and sponsorship

Nil.

Conflicts of interest

There are no conflicts of interest.

 
  References Top

1.
Szymanska J, Sitkowska J, Dutkiewicz J. Microbial contamination of dental unit waterlines. Ann Agric Environ Med 2008;15:173-9.  Back to cited text no. 1
    
2.
O'Donnell MJ, Boyle MA, Russell RJ, Coleman DC. Management of dental unit waterline biofilms in the 21st century. Future Microbiol 2011;6:1209-26.  Back to cited text no. 2
    
3.
Kohn WG, Collins AS, Cleveland JL, Harte JA, Eklund KJ, Malvitz DM; Centers for Disease Control and Prevention (CDC). Guidelines for infection control in dental health-care settings-2003. MMWR Recomm Rep 2003;52:1-61.  Back to cited text no. 3
    
4.
Atlas RM, Williams JF, Huntington MK. Legionella contamination of dental-unit waters. Appl Environ Microbiol 1995;61:1208-13.  Back to cited text no. 4
[PUBMED]    
5.
Ma'ayeh SY, Al-Hiyasat AS, Hindiyeh MY, Khader YS. Legionella pneumophila contamination of a dental unit water line system in a dental teaching centre. Int J Dent Hyg 2008;6:48-55.  Back to cited text no. 5
    
6.
Shepherd PA, Shojaei MA, Eleazer PD, Van Stewart A, Staat RH. Clearance of biofilms from dental unit waterlines through the use of hydroperoxide ion-phase transfer catalysts. Quintessence Int 2001;32:755-61.  Back to cited text no. 6
[PUBMED]    
7.
Petti S, Moroni C, Messano GA, Polimeni A. Detection of oral streptococci in dental unit water lines after therapy with air turbine handpiece: Biological fluid retraction more frequent than expected. Future Microbiol 2013;8:413-21.  Back to cited text no. 7
[PUBMED]    
8.
Berlutti F, Testarelli L, Vaia F, De Luca M, Dolci G. Efficacy of anti-retraction devices in preventing bacterial contamination of dental unit water lines. J Dent 2003;31:105-10.  Back to cited text no. 8
[PUBMED]    
9.
Tuttlebee CM, O'Donnell MJ, Keane CT, Russell RJ, Sullivan DJ, Falkiner F, et al. Effective control of dental chair unit waterline biofilm and marked reduction of bacterial contamination of output water using two peroxide-based disinfectants. J Hosp Infect 2002;52:192-205.  Back to cited text no. 9
    
10.
Anonymous. Legionnaires' disease: A guide for employers. UK: Health and Safety Executive; 2001.  Back to cited text no. 10
    
11.
Walker JT, Bradshaw DJ, Fulford MR, Marsh PD. Microbiological evaluation of a range of disinfectant products to control mixed-species biofilm contamination in a laboratory model of a dental unit water system. Appl Environ Microbiol 2003;69:3327-32.  Back to cited text no. 11
[PUBMED]    
12.
Majtán V, Majtánová L. Effect of quaternary ammonium salts on energy-yielding and energy-requiring processes in Salmonella typhimurium. Arzneimittelforschung 1995;45:1021-3.  Back to cited text no. 12
    
13.
Majtán V, Majtánová L, Hostacká A, Hybenová D, Mlynarcik D. Effect of quaternary ammonium salts and amine oxides on Pseudomonas aeruginosa. Microbios 1995;84:41-51.  Back to cited text no. 13
    
14.
Csiba I, Devínsky F, Masárová L, Lacko I. Organic ammonium salts. XLI. Preparation, aggregation properties and antimicrobial activity of N-[2-(10-undecenoyloxy)ethyl]-N, N, N-alkyldimethylammonium bromides. Cesk Farm 1991;40:148-51.  Back to cited text no. 14
    
15.
Kopecká-Leitmanová A, Devínsky F, Mlynarcik D, Lacko I. Interaction of amine oxides and quaternary ammonium salts with membrane and membrane-associated processes in E. coli cells: Mode of action. Drug Metabol Drug Interact 1989;7:29-51.  Back to cited text no. 15
    
16.
Fraise AP. Choosing disinfectants. J Hosp Infect 1999;43:255-64.  Back to cited text no. 16
[PUBMED]    
17.
Russell AD. Assessment of sporicidal efficacy. J Int Biodeterior Biodegradation 1998;4:281-7.  Back to cited text no. 17
    
18.
Aksen F, Kaya A, Akpolat N. Comparison of bactericidal activity of microwave, ultraviolet and disinfectant solutions on some bacterial strains. Biotechnol Biotechnol Equip 2004;18:122-27.  Back to cited text no. 18
    
19.
Holton J, Shetty N, McDonald V. Efficacy of 'Nu-Cidex' (0.35% peracetic acid) against mycobacteria and cryptosporidia. J Hosp Infect 1995;31:235-7.  Back to cited text no. 19
[PUBMED]    
20.
Liaqat I, Sabri AN. Effects of biocides on bio-film bacteria from dental unit waterline biofilms. Curr Microbiol 2008;6:1220-7.  Back to cited text no. 20
    
21.
Staniszewska ME, Rodowald R, Jakimiak B. Disinfection action of disinfectants. Infection 2006;6:14-7.  Back to cited text no. 21
    



 
 
    Tables

  [Table 1], [Table 2]



 

Top
 
 
  Search
 
Similar in PUBMED
   Search Pubmed for
   Search in Google Scholar for
 Related articles
Access Statistics
Email Alert *
Add to My List *
* Registration required (free)

 
  In this article
Abstract
Introduction
Methodology
Results
Discussion
Conclusion
References
Article Tables

 Article Access Statistics
    Viewed21    
    Printed0    
    Emailed0    
    PDF Downloaded9    
    Comments [Add]    

Recommend this journal


[TAG2]
[TAG3]
[TAG4]